46 Cards in this Set
- Front
- Back
widening of the sound beam is demonstrated in the
|
fraunhofer zone
|
|
weak beams emitted from a linear sequenced array transducer are termed
|
grating lobes
|
|
the resonant frequency of a pulse wave is determined by the
|
thickness and propagation speed of the element
|
|
heat sterilization is not recommended for diagnostic transducers because
|
the piezoelectric properties will be lost
|
|
which of the following compontnts is unnecessary in the construction of a continuous wave transducer |
damping layer
|
|
lateral resolution is determined by the |
beam width
|
|
what is the axial resolution in soft tissue when using a 5 MHz frequency with a two cycle pulse and an element thickness of 0.5mm
|
0.3 mm
|
|
what is the operating frequency of a two cycle pulse with an element thickness of 0.2 mm and a propagation speed of 4 mm / us
|
10 MHz
|
|
which of the following principles do diagnostic ultrasound transducers operate? snell's law ALARA principle Hugens principle piezoelectric effect |
piezoelectric effect
|
|
reducing the impedance difference between the crystal and the skin is the primary function of what transducer component |
matching layer
|
|
a sound beam demonstrates the most uniform intensity in the |
far field
|
|
what is the near zone length of a 6 mm 5MHz transducer
|
30mm
|
|
which of the following changes will improve temporal resolution increase the beam width increase focal zone depth decrease the depth decrease in dynamic range |
decrease the depth
|
|
The near zone length of a 3 mm 10 MHz transducer is |
15mm
|
|
which of the following transducers elements has the longest focal length 7mm, 2.5 MHz 2mm, 7.5 MHz 5mm, 10 MHz 3mm, 15 MHz |
5mm, 10MHz
|
|
the distance from the face of a focused transducer to the point of spatial peak intensity is termed the
|
focal length |
|
Irregular excitation of the elements in an array to reduce grating lobes describes
|
apodization
|
|
diameter of the transducer describes the
|
diameter of the focus
|
|
the impedance of the damping layer is
|
similar to the element's
|
|
the purpose of backing material in the transducer is to
|
reduce the number of cycles in a pulse
|
|
vector array is a type of transducer
|
operation
|
|
focusing of the sound beam is directly related to
|
lateral resolution
|
|
steering of the sound beam is accomplished by
|
altering the excitation of the active elements
|
|
exceeding the curie point of a transducer wll result in
|
loss of all piezoelectric properties
|
|
an active element with a thickness of 0.8 mm and a propagation speed of 4 mm/us will have an operating frequency of
|
2.5MHz
|
|
what is the thickness of the crystal with an operating frequency of 5 MHz and a propagation speed of 4 mm/us? |
0.4 mm
|
|
what is the axial resolution in soft tissue when using a 15 MHz frequency with a two cycle pulse |
0.1 mm
|
|
which of the following transducers operates by applying voltage pulses to groups of linear elements in succession |
linear sequenced array
|
|
focusing of the sound beam is only accomplished within the
|
near field
|
|
what is the diameter of the sound beam at one near zone length when using a 6mm 2.5 MHz transducer
|
3mm
|
|
the ability to distinguish two structures in a path perpendicular to the sound beam describes
|
azimuthal resolution
|
|
the most common piezoelectric material used in diagnostic ultrasound transducers is
|
lead zirconate titanate
|
|
subdicing the elements in diagnostic ultrasound transducers is used to |
reduce grating lobes
|
|
diagnostic frequencies range between
|
2 and 20 MHz
|
|
which of the following changes will improve axial resolution
|
increase in transducer frequency
|
|
during transmission diagnostic ultrasound transducers convert
|
electrical energy into thermal energy
|
|
which of the following states "some materials produce a voltage when distorted by an applied pressure"
|
piezoelectric principle
|
|
which of the following is a negative effect of using a damping material in the transducer assembly
|
reduced sensitivity
|
|
how many cycles pre pulse are typically used in doppler imaging
|
5-30
|
|
the sonographer can determine the depth of the focal zone when using a transducer with an
|
electronic focus
|
|
the impedance of the matching layer is
|
less than the impedance of the crystal
|
|
which of the following transducers displays a trapezoidal image? convex vector annular curvilinear |
vector |
|
which of the following differentiates similar or dissimilar tissues? axial resolution lateral resolution contrast resolution temporal resolution |
contrast resolution
|
|
section thickness is related to the
|
beam width
|
|
operating frequency of a transducer is directly related to the
|
propagation speed of the element
|
|
a focused 10 mm 15 MHz transducer will demonstrate a focal diameter of
|
5 mm
|