What is the inverse square law?
|
If distance doubles the intensity of radiation decreases four fold and inversely if distance is cut in half, intensity increases fourfold.
what effects its validity? |
scatter radiation, has to be a point source
|
|
What is the inverse square law equation?
|
I 1 / I 2 = d 2 squared/d 1 squared
|
|
|
A gamma source produces a reading of 150 mRem/hr at 3m. At what distance will the meter read 500 mRem/hr?
|
1.64m
A syringe containing 5 mCi of Tc99m reads 25mR/hr at 1 meter. What is the dose rate at the surface of the syringe, assuming the syringe is 2cm? |
250,000 mR/hr
|
|
What is the allowable dose to the public?
|
2 mR/week or 100 mR/yr or no single dose of 2 mR/hr
|
|
|
What are the 2 areas NM depts are divided into?
|
Restricted and Non-restricted
give some examples |
Restricted: hot lab, radioactive storage areas
Non-restricted: waiting or scanning area, office, etc |
|
What are the 6 factors to consider for shielding radioactive material?
|
1. occupancy factor
2. distance from the source 3. how long is the source present 4. area restricted or non-restricted 5. type radiation sourced emits 6. existing shielding |
|
|
How is an image created with a CT scanner?
|
the average linear attenuation coefficient change in intensity between tube and detector
|
|
|
What is the average linear attenuation coefficient for water, air and bone?
and how is the CT# arrived at? |
water=0; air=-1000; bone=1000
CT# is calculated by µtissue - µwater/ µwater x 1000 |
|
|
What are the advantages of PET/CT fusion imaging?
|
small time between PET and CT, body does not change
anatomy is better visualized with CT rather than PET alone greatly improves ability to locate lesions and other patholgy |
|
|
Why is PET/CT not perfect?
|
CT is much faster than PET so motion artifacts can occur, especially with lungs
difference in matrix sizes |
|