Selection of the incorrect body part from the examination menu may result in _____
image misinterpretation
kilovoltage peak should be chosen ____
for the type and amount of contrast desired
the k-edge of phosphor imaging plates ranges from ____ keV
30 to 50
the range of kilovoltage peak that can be used with CR is ____ KVp
45 to 120
True or False: less kilovoltage peak is needed with conventional radiographic screens than is necessary for CR systems
False
mas is selected for the _____
number of electrons needed
when insufficient light is produced by the imaging plate phosphor, the image will _____
be grainy
quantum mottle is caused by ____
insufficient light
True or False: backscatter from the cassette-detector influences the amount of mAs necessary to produce the image
True
using an imaging plate that is much larger than the part being examined will ______ resolution
decrease
True or False: using the wrong imaging plate size may result in a matrix with very small pixels
False
the higher the number of pixels in a matrix, the _______ the image resolution
higher
true or False: grid lines projected onto the imaging plate will not interfere with the image
False
the moire artifact results when grid lines and scanning lasers are ______
parallel
Grid use for CR examinations is more critical than in film/screen radiography because CR _____
imaging plates record more densities
the number of grid lines per centimeter or lines per inch describes grid ____
frequency
Grid ratio describes the ____
relationship between the height of the lead strips to the interspace
which of the following requires more critical beam centering?
-parallel grids
-focused grids
-low-ration grids
-low-frequency girds
focused grids
collimation to a smaller area _____
decreases compton interactions
true or False: it is acceptable to replace collimation with shuttering
false
True or False: images should be marked only with computerized markers
False
the exposure indicator number for Fuji systems is the ____ number
S
the exposure indicator number is an _____
indication of radiation absorbed by the imaging plate
true or False: with Fuji exposure indicators, the lower the S number, the higher the exposure
true
What manufacturer uses exposure indicator numbers in the thousands?
Kodak
As kodak exposure indicator numbers increase by 300, exposure
increases by a factor of 2
an S number of 400 in a Fuji system ____ an S number of 2000
is half the exposure of
in a Fuji system, each change of exposure number of 200 is equal to a change by a factor of ___
2
in a ll systems, exposures outside the range indicate ____
both overexposure and underexposure
a histogram is____
a graphic representation of x-ray exposure
dark line artifacts can be caused by ____
scatter through the back of the cassette
horizontal white lines may be caused by ____
dirt on the light guide in the reader
to avoid the moire grid artifact, the grid must be oriented ____ to the plate readers laser scan direction
perpendicular
fine white lines appearing on a hard copy of the image may be caused by _____
debris on the mirror in the laser printer