True or False
If input one is more negative than input 2 the deflection of the waveform will be up |
True
|
|
Impedance is defined as:
|
the resistance against alternating current flow
|
|
A bipolar montage follows which rules?
|
combines serial pairs of electrodes into a chain, in a chain of electrodes input 2 of one channel is the same electrode as input 1 of the next, left side electrodes appear first followed by the right side
|
|
What is the amplitude (uV) of activity that measured 12 mm in height, if a 50 uV calibration signal measured 6 mm?
|
100 uV
|
|
nasopharyngeal electrodes are placed where:
|
through the nasal passage
|
|
If the vertical scale (sensitivity) is 5 uV/mm, then a 60uV will measure:
|
12 mm
(60uV/5uV/mm) |
|
Best montage to measure the amplitude of a sharp wave?
|
A2 ear reference montage
|
|
A faster sweep speed (60mm/sec) allows:
|
easier measurements of faster frequencies, easier ID of 60 hz artifact, better ID of degree of synchrony between 2 hemispheres.
|
|
The use of 60hz notch filter does what:
|
eliminates 60 hz activity only
|
|
The average reference recording is best used for:
|
localize focal activity
|
|
Sensitivity is conventionally defined by:
|
ratio of voltage to height
|
|
What type of montage combines activity from many electrodes to form a reference?
|
Average reference
|
|
which type of montage most gives the true amplitude EEG activity at a given electrode?
|
non-cephalic reference
|
|
The time constant of .1 second is aprox. equivalent to what low filter setting?
|
1 Hz
|
|
Common mode rejection ratio is closely related to ?
|
A balanced amplifier
|
|
what is the least invasive electrode application?
|
subdermal
|
|
A type of amplifier that amplifies the differences in voltage between 2 inputs?
|
differential amplifier
|
|
electrode "pop" is induce by what?
|
poorly chlorided electrodes
|
|
The time constant is determined by?
|
resistance x capacitance
|
|
The laplacian montage is often used for?
|
topographic EEG mapping
|
|
time scale of 3 cm/sec, what is the duration in mm for a 10 hz wave?
|
3 mm
|
|
The average reference is created by :
|
averaging the activity from all or (most) electrodes used.
|
|
High pas filter is:
|
eqivelant to low frequency filter
|
|
bipolar montage is best used to:
|
localize focal activity
|
|
True or false
The bio cal demonstrates the integrity of the system from patient to instrument. |
True
|
|
Transverse bipolar monitoring is more accurate than longitudinal montage in localizing the activity:
|
spreading from temporal to parasagittal region.
|
|
What is the duration of a wave measuing 5 mm, when the 20 mm calibration signal is equal to 200 msec?
|
50 msec
(10 x 5mm = 50) |
|
Which seizure type (ictal event) least likely shows abnormality?
|
?
|