Sterile items that are protected from the environment have been proven to remain sterile |
Indefinitely |
|
The period of time during which a sterile item is considered safe to use is known as |
shelf life |
|
Which of the following factors is a reason for a manufacturer to assign an expiration date to a sterile medical device?
A. Product labeling B. Product degradation C. The cost of the product D. The type of sterilization method |
B. Product degradation |
|
A package must be visually inspected for integrity |
every time it is handled |
|
Sterile trays that might be subjected to storage challenges should be |
placed in sterility maintenance cover |
|
Stock rotation and frequent evaluation of stock levels will minimize the need for reprocessing and prevent |
outdates |
|
What should be done if a reusable sterile basin set is dropped on the floor? |
Send it to be reprocessed |
|
The sterile storage area should be located |
adjacent to the sterilization area |
|
Humidity can be measured by using a |
hygrometer |
|
The minimum number of air exchanges per hour in the sterile storage areas should be |
4 |
|
The sterile storage area should be maintained at |
positive pressure |
|
The relative humidity in the sterile storage area should not exceed |
70% |
|
The temperature in the sterile storage area should not exceed |
75 F (24 C) |
|
It is important to control the humidity in the sterile storage area in order to prevent |
the growth of microorganisms |
|
Shelving in the sterile storage area must be kept at what distance from the outside walls? |
2 inches |
|
Shipping cartons are not acceptable storage containers in the sterile storage area because they |
harbor soils |
|
Rotating stock to ensure that the oldest supplies are used first is known as |
FIFO (first in first out) |