17 Cards in this Set

  • Front
  • Back

Sterile items that are protected from the environment have been proven to remain sterile

Indefinitely

The period of time during which a sterile item is considered safe to use is known as

shelf life

Which of the following factors is a reason for a manufacturer to assign an expiration date to a sterile medical device?



A. Product labeling


B. Product degradation


C. The cost of the product


D. The type of sterilization method

B. Product degradation

A package must be visually inspected for integrity

every time it is handled

Sterile trays that might be subjected to storage challenges should be

placed in sterility maintenance cover

Stock rotation and frequent evaluation of stock levels will minimize the need for reprocessing and prevent

outdates

What should be done if a reusable sterile basin set is dropped on the floor?

Send it to be reprocessed

The sterile storage area should be located

adjacent to the sterilization area

Humidity can be measured by using a

hygrometer

The minimum number of air exchanges per hour in the sterile storage areas should be

4

The sterile storage area should be maintained at

positive pressure

The relative humidity in the sterile storage area should not exceed

70%

The temperature in the sterile storage area should not exceed

75 F (24 C)

It is important to control the humidity in the sterile storage area in order to prevent

the growth of microorganisms

Shelving in the sterile storage area must be kept at what distance from the outside walls?

2 inches

Shipping cartons are not acceptable storage containers in the sterile storage area because they

harbor soils

Rotating stock to ensure that the oldest supplies are used first is known as

FIFO (first in first out)