The high voltage that is required by an x-ray machine is produced by the ____________.

transformer

A transformer works on the principle of ____________.

mutual induction

The primary coil is the ________ side of the transformer.

input

The secondary coil is the ____________ side of the transformer.

output

The transformer law states that the emf induced in the secondary coil is to the emf in the primary coil, as the _____________ in the secondary coil is the ____________ in the primary coil.

number of turns

A ______________ transformer has more turns in the secondary coil than in the primary coil, and therefore _____________ is created in the secondary coil than is supplied to the primary coil.

step up; higher voltage

A ___________ transformer has fewer turns in the secondary coil than in the primary coil and therefore _________ is created in the secondary coil than was supplied to the primary coil.

step down; less voltage

According to the law of conservation of energy, there can be no more ________ coming out of the transformer than was put in.

energy

A step up transformer increases ___________ and decreases ______________.

voltages; amperage

A step down transformer decreases ________ and increase ___________.

voltage; amperage

In a diagnostic x-ray machine, the transformer provides the_____________ necessary for the electrons to be driven towards the anode with enough speed to create an x-ray photons.

high potential difference.

The step up transformer of an x-ray machine changes voltage into ____________.

kV

The step up transformer of an x-ray machine changes amperage into _________.

mA

The leakage of magnetic flux in the air at the ends of the cores of a transformer is known as _____________.

flux leakage

The ratio of power output to power input of a transformer is known as ___________.

efficiency

The loss of electrical power due to the resistance of the resistance of the coils of a transformer is known as __________.

copper loss

In a step up transformer, the primary winding carries a larger current so the wire must be ________ as compared to the secondary coil which carries a smaller current and therefore the wire may be ___________.

thicker; smaller

The swirling current created in the core which is induced by the magnetic flux is called an __________.

eddy current

The power loss of a transformer due to the energy necessary to overcome the opposition to change, exhibited by ferromagnetic materials, every time the magnetic field produced by the primary winding changes polarity is known as ____________.

hysteresis

The ___________ varies input voltage to the transformer in order to control output voltage and therefore functions as the kV selector of an x-ray machine.

autotransformer

The autotransformer is connected between the ____________ and the __________ side of the the transformer.

incoming power source; primary coil

The autotransformer operated on the principle of _________.

self induction

The secondary voltage of the autotransformer is applied to the ________ side of the main step up transformer.

primary

The _________ circuit of the x-ray tube carries the current needed to heat the filament.

filament

The _________ circuit of the x-ray tube carries the current passing between the cathode and anode.

tube or main

A ________ is a coil with a core whose position is able to be adjusted, which operates on the principle of self induction and was once used to regulate filament current in the x-ray tube.

Choke Coil

A ________ is a variable resistor designed to regulate filament current

rhetostat

Two coils wound around a square iron core.

closed core transformer

A single winding that serves as both the primary and secondary coil, with an adjustable number of windings on the secondary side.

autotransformer

Two insulated coils of wire with an iron core side by side.

open core transformer

Primary and secondary coils wound around the central section of a core consisting of sheets of silicon steel with two rectangular cutouts.

shell type transformer

Two insulated coils of wire side by side.

air core transformer

Which type of transformers have most flux leakage?

open and air core transformers

Which type of transformer is most commonly used in x-ray equipment?

closed core transformer

Is a transformer considered to be efficient?

yes

How can copper loss be reduced?

by giving the side with the greater current a thicker wire

How can eddy currents be reduced?

it is reduced by laminated silicon steel core