Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
42 Cards in this Set
- Front
- Back
The process which all forms pf microbial life, including spores, are destroyed |
sterilization |
|
A gauge on the main steam line, usually located at the back of the sterilizer, tells you the steam pressure |
as it is entering the jacket |
|
Minimum exposure time for wrapped goods in a dynamic-air-removal steam cycle at 275* F (135* C) and 28 - 30 pounds per square inch (psi) |
3 minutes |
|
Immediate-use steam sterilization may be indicated when |
a one-of-a-kind instrument is contaminated |
|
Best describes a "wet load" |
multiple packages in the same load found with excess moisture |
|
For sterilization to occur, steam MUST |
make direct contact with all surfaces |
|
This action should be taken if a sterilized load is determined to be wet |
reprocess the entire load |
|
Can occur if packages are moist at the end of the steam cycle |
microorganisms can wick into the package |
|
For sterilization, paper-plastic peel packages should be placed |
on their edge |
|
In a mixed load of textile and metal goods, the textile packs should be placed |
on the top rack |
|
For steam sterilization, basin sets should be positioned |
tilted on their sides |
|
Stacking of rigid containers in a steam sterilizer is permitted when |
the manufacturer provides technical data showing that it is acceptable |
|
Is used to detect air leaks or insufficient air removal |
Bowie-Dick test |
|
One reason for adding chemicals to the boiler feed-water is to |
minimize contaminates |
|
Which monitoring method verifies that conditions within the load were adequate to kill bacterial spore |
biological |
|
What is the CORRECT action to take if an item needs to be corrected or removed from the sterilization load record |
draw a line through the item and initial |
|
The Bowie-Dick test is considered which type of monitoring process |
chemical |
|
A class 5 indicator is called a |
integrating indicator |
|
Class 6 indicators are also called |
emulating indicators |
|
What should be done if the Bowie-Dick test fails the first time |
report to the supervisor for further instruction |
|
Is the RECOMMENDED sterilizer setting for the Bowie-Dick test |
3 1/2 to 4 minutes at 270*F (132*C) dynamic-air- removal |
|
Generally, if a biological test is performed daily, how often is it necessary to incubate a control biological indicatir |
once a day and if a new lot number is used |
|
Routine biological testing of steam sterilizers should be performed no less then |
weekly |
|
All loads containing implantable devices should be |
monitored with a biological indicator |
|
For testing of rigid sterilization containers, the biological indicators should be placed |
as recommended by the manufacturer of the container |
|
Sterile packs are most vulnerable to contamination |
when the package is still hot |
|
Items being unloaded from the steam sterilizer should be handled with |
clean hands |
|
A loaner instrument has been contaminated during a procedure and the OR wants to sterilize it in the immediate-use steam sterilizer. What would be the first step to take |
consult the manufacturer's written instructions for approved sterilization methods and parameters |
|
Best defines "superheating" of steam |
heat added to dry steam in the absence of water |
|
Sterilization method that would require aeration after processing |
Ethylene oxide |
|
The basic parameters of gas sterilization in a 100% ethylene oxide sterilizer are |
concentration, exposure time, humidity and temperature |
|
Sterilizer that requires only a special electrical outlet for installation |
Low-temperature gas plasma |
|
All devices processed by low-temperature gas plasma must be |
clean and dry |
|
Action that should be taken when a positive BI demonstrates a sterilization process, failure, and the failure is the result of an operator error |
the error should be corrected and the load reprocessed |
|
Generally, the ethylene oxide process challenge device (Biological Test Pack) is placed in the |
geometric center of the load |
|
The AAMI recommends that rigid sterilization containers be biologically tested before purchase and annually to verify that the |
sterilizer is able to sterilize the contents |
|
In qualification testing of steam sterilizers, to establish a baseline for the sterilizer's performance, when should the BI process challenge device (PCD) test be performed |
before the Bowie-Dick tests |
|
Should not be processed in the low-temperature gas plasma sterilizers |
gauze sponges |
|
The parameters for low-temperature gas plasma sterilization are |
time, temperature, and hydrogen peroxide gas plasma |
|
Is compatible with ozone sterilization |
woven textiles |
|
Microorganisms that are used to biologically monitor the ozone sterilizer |
Geobacillus stearothermophilus |
|
The parameters of ozone sterilization are |
concentration, temperature and time |